QUANTITATIVE MEASUREMENT OF ENZYME ACTIVITY
The invention relates to a novel method for the quantitative measurement of enzyme activity in eluants from dried bloodspots (DBS) or whole blood samples and in particular is suitable for screening deficiencies such as Glucose-6- Phospate Dehydrogenase deficiency in newborns.
Glucose-6-Phosphate dehydrogenase (G-6-PD) is a cytoplasmic enzyme, which is distributed in all cells. It catalyses the first step in the hexose monophosphate pathway producing NADPH as shown below.
G6PD
Reductive biosyπthβtic reactions.
The main metabolic role of G-6-PD in red cells is the defense against oxidizing agents, epitomized by hydrogen peroxide. NADPH, a product of the G-6-PD reaction, is both the hydrogen donor for regeneration of reduced glutathione and a ligand for catalase (see text). GSSGR = glutathione reductase; GSHPX = glutathione peroxidase; G-6-P = glucose-6-phosphate; 6PG = phosphoglucnnate.
This coenzyme is required as a hydrogen donor for reactions of various biochemical pathways as well as for the stability of catalase and the preservation and regeneration of the reduced form of glutathione. Catalase and glutathione are both essential for the detoxification of hydrogen peroxide, therefore the defence of the cells against H2O2 is heavily dependent on G-6-PD. The red cells are very sensitive to oxidative damage and lack other NADPH producing enzymes. The defence against oxidising agents, (epitomised by H2O2) is mainly realised by glutathione, which converts H2O2 to H2O stoichiometrically via glutathione peroxidase. NADPH is the hydrogen donor for the regeneration of reduced glutathione. An alternative pathway for H2O2 detoxification is via catalase, but this route is regarded as ineffective under normal conditions because of the lower affinity of catalase for H2O2 compared to that of glutathione peroxidase.
G-6-PD deficiency is the most commonly known enzymopathy with around 400 million people affected worldwide. The prevalence ranges from 5% to 25% in the endemic areas such as Africa, the Middle East, Asia, the Mediterranean and Papua New Guinea but the highest incidence of 65% is found in Kurdish Jews. Incidences ranging from 0.5 to 6.9% have also been reported in North and South America. Around 400 mutations have been reported so far.
Clinical manifestations associated with G-6-PD deficiency include:
1). Drug induced hemolysis - certain antimalarials, suiphonamides, sulfones and other drugs or chemicals are associated with significant hemolysis in subjects;
2). Infection induced hemolysis - numerous bacterial, viral and rickettsial infections have precipitated hemolysis, but the most important are infectious hepatitis, pneumonia and typhoid fever;
3). Favism - sudden onset of acute haemolytic anaemia within 24 to 48 hours of ingesting fava beans;
4). Neonatal jaundice - jaundice usually appears by 1 to 4 days of age 5). Chronic nonsperocytic haemolytic anaemia.
Such deficiencies have usually been measured by semi-quantitative assays such as those described by Beutler et al where he measured the enzyme activity in dried bloodspots (DBS). The principle of the methods described by Beutler is to take a sample by hole punch from the DBS, elute it with a haemolysing solution and incubate the sample from this with a reaction mixture at 37°C.
Subsequently adding an acid solution to the mixture stops the reaction and an aliquot of said reaction mixture is then spotted onto filtration paper.
The NADPH formed by the above reaction is then visualised using a fluorimetric method by looking at the samples under ultra-violet light in a dark room. This gives a semi-quantitative interpretation, comparing the fluorescence of the sample with controls and the results are classified normal, intermediate or deficient. It should however be noted that such a test must only discriminate between normal and abnormal. Therefore if a limit is chosen (i.e. anything lower than 3.5 U/g Hb is to be regarded as not normal) the test only needs to make sure that it will not give results when the activity is lower than that
(acceptance tests), or alternatively, will give results if the activity is lower than that (rejection tests). They do not give results in the whole range. Other problems with this method include that there is frequently interference from other enzymes producing NADPH, the method can only work at temperatures of 37°C and spontaneous fluorescence can cause misclassification of the sample.
Some commercial companies have modified the above method for use with automated fluorimetric platereaders ( Isolab, Labsystems ). Such instruments produce a quantitative result by measuring the amount of fluorescence and calculating the enzyme activity from this. There are also problems with these automated fluorimetric platereaders as they only work at temperatures of 37°C, there is a problem of interference from other enzymes producing NADPH, spontaneous fluorescence can cause misclassification of the sample and the quantitative result obtained therefrom is not reliable because it is influenced by the amount of haemoglobin eluted from the DBS. This can lead to misclassification of the sample.
The object of the present invention is to obviate or mitigate the problems of these prior art methods to provide an improved method capable of producing reliable results which can be used for analysing both DBS and whole blood samples.
According to the present invention there is provided a quantitative assay for an analyte of interest in blood samples, comprising (a) the preparation of a sample for analysis;
(b) a kinetic photometric measurement of the sample, being periodic measurements recorded under a controlled temperature, at a selected wavelength in the range of about 330-360nm; said assay being characterised by
(c) a photometric measurement of the sample at a wavelength in the range of 400-420nm or 535-545nm or 565-580nm; (d) conducting steps equivalent to (b) and (c) above upon a control; and
(e) calculating the result of the analysis by solving the following δODsλl/min/δODcλl/min x Control Value = Sample Value (Activity in U/g Hb) ODsλ2/ODcλ2 wherein δODs is the change in optical density (per minute) for the sample, λl is the wavelength at which optical density is measured under kinetic mode, δODc is the change in optical density (per minute) of a control measured under kinetic mode conditions at a particular wavelength (λl), ODs is the optical density for the sample measured once at a particular wavelength (λ2), and ODc is the optical density of the control measured once at that particular wavelength (λ2), λ2 is the wavelength at which a reading of Heamoglobin contained can be taken, the control value being determined by the change in optical density (per minute) of the control measured under kinetic mode at λl divided by the optical density of the control at λ2 i.e. δODcλl/mim/ODcλ2.
Preferably the sample preparation is by a pre-incubation of a sample at a selected temperature, usually within the range of from about 20°C to about 40°C, preferably at a temperature of about 22-25°C, or about 22-30°C, or about 37°C, for a period dependent upon the incubation temperature and typically not exceeding about 10 minutes at temperatures around 22-25°C, about 5 minutes for temperatures around 22-30°C and about 2 minutes for a temperature of about 37°C.
The invention has particular utility for analytes of interest to be found in dried blood stains (DBS) but is surprisingly effective also for whole blood, including red blood cells in solution e.g. saline. The assay is particularly useful in quantitative measurement of enzyme activity in such samples, and enables assessment of the haemoglobin content of these samples and provides results in terms of Units per gram haemoglobin.
The sample preparation may be such that the sample is obtained from a dried blood stain on a physical support or carrier such as filter paper which can be hole punched to obtain a suitable sample "spot", this spot may be then incubated with haemolysing solution, and an aliquot of that may subsequently be incubated with substrate, coenzyme and oxidising agent. Elution buffer and dilution buffer may be present in the sample when presented for analysis.
The method will now be further described below with reference to the examples described below.
The present invention provides a method for the quantification of enzyme activity in eluates from DBS on filter paper or from whole blood samples in microtiter plates. In particular the method includes an assay utilising G-6-PD, which in the presence of NADP, catalyses the oxidation of G-6-P to 6- phosphogiuconate. The NADPH produced is measured coiorimetrically at 340nm in kinetic mode and the results are calculated by evaluating the increase in optical density (OD) per minute (slope) for unknowns against the slope for a Standard with known G-6-PD activity. However, it will be understood by those in the art that such a method as described can also be used to measure many other such deficiencies and its application is not solely limited to measuring G-6- PD deficiencies.
The first step is carried out by taking a sample from a DBS on filter paper using a hole punch and incubating this spot with a haemolysing solution., a sample of which is then incubated with a reaction mixture. Alternatively whole blood samples can be measured in micro titre plates by addition of the reagents to the wells in the micro titre plate which contain the whole blood.
The reaction mixture contains: Substrate - Glucose-6-Phosphate
Coenzyme - β-NADP Oxidising agent - L-Glutathione
Dilution Buffer - Phosphate buffered saline rated at pH 7.9 with the addition of Na2HPO4 (Sodium Monohydrogen Phosphate) and Sodium
Azide as preservative (0.12% w/v)). Elution buffer - Hypotonic solution which is a mixture of 0.01M Na2HPO4 and 0.01M NaH2PO4 rated at a pH of 7.15 and containing sodium azide as a preservative (0.12%w/v).
A pre-incubation of the sample in the reagent mixture is then allowed. In assays performed at 22-25°C the incubation time is 10 minutes, at 22-30°C it is 5 minutes, and at 37°C it is 2 minutes. The incubation period is then followed by photometric measurements at 340nm at 1 minute intervals up to a maximum of 25 measurements after which the reaction rate is expressed in mOD/min.
The second step involves a single reading of the same reaction mixture at 400- 420nm or 535-545nm or 565-580nm. This reading is representative of the amount of haemoglobin eluted from the filter paper. Alternatively a sample of the same eluate can be mixed with the Dilution Buffer using the same volumes as in step one, followed by a single reading at 400- 420nm or 535-545nm or 565-580nm.
The result of the first step ( mOD/min ) is divided by the result of the second step (OD) and multiplied by a FACTOR X to express the final test result in Units per gram Haemoglobin. The Factor is calculated using a standardised solution with known Enzyme Activity and Haemoglobin content.
Procedure
The samples are collected from the patient by standard procedures. The reagents are warmed up slowly to 18-25°C prior to use. 14ml of the oxidizing agent is mixed with 14ml of dilution buffer, 14ml of substrate is mixed with
buffer. The reaction mixture is prepared by mixing equal volumes of the above dilutions. Approximately 25μl of each component is required for each test.
Procedure for carrying out the test with micro titration plates
1. Punch blood spots approximately 3/16 inch in diameter in U bottomed micro titre plate. Use position Al as Normal Control and use 2 wells for the
Intermediate and Deficient Controls.
2. Add 75μl of Elution buffer to each well.
3. Place micro titre plate on orbital shaker for 30 minutes at room temperature. 4. During the elution prepare Reagent Mixture. (If performed at 30°C the reagent and a flat bottomed micro titre plate need to be warmed to 30°C).
5. Transfer 15μl of the eluate from each well to the corresponding well of the new flat bottom microtiter plate.
6. Add 75μl of the reagent mixture to each well. 7. Place the plate on an orbital shaker for 1 minute.
8. Place the plate in the incubator and wait until (a)10 minutes have passed after reagent mixture has been added for temperature range 22-25°C or (b) 5 minutes have passed after reagent mixture has been added for temperature range 25-30°C. 9. Read the plate in a plate reader at 340nm for 15 minutes with 60 second intervals.
10. After the readings are completed, change the program of the reader to a single reading at 405nm and read the plate (containing the same mixture) again. Procedure for carrying out the test with Cobas Mira (37°C)
1. Punch blood spots 2 by 1/8 inch in a sample cup, leave caps open.
2. Add 150μl of elution buffer to each well.
3. Place sample rack on an orbital shaker for 30 minutes.
4. During the elution, reconstitute reagents if needed and prepare reagent mixture (1 test uses 150μl of reagent mixture).
5. Close sample cups and place sample rack on the instrument.
6. Load reagent mix.
7. Insert empty cuvettes, and run the test.
Procedure for carrying out the test with whole blood specimens 1. Pipette 10μl of whole blood into a sample cup
2. Add 150μl of Elution buffer
3. Close caps and mix gently
4. Load reagent mix
5. Insert empty cuvettes, and run test. Hb Protocol Cobas Mira
Run the test using the same sample as used for the G-6-PD activity assay. There is sufficient sample in the cup for a second test. The reagent mix becomes the dilution buffer in this case and instrument settings should be set at 405nm to run the test.
The G-6-PD activity of the sample is directly comparable to that of the control (OD control 340nm/min). Since the control's activity is already given in Units/g Hb (U/g Hb) the sample's activity can readily be expressed in the same units. The equation which gives results in U/g Hb is given below.
δODSample340nm/min / δODcontrθ|340nm/nnin X Control Value = Sample Activity in U/g Hb
ODsa p|e405nm / ODControl405nm
(OD sample at 405nm (whereby 405 can be substituted by any wavelength in the range 400-420nm or 535-545nm or 565-580nm))
A kit is also provided for carrying out the method which generally comprises 8 rubber stoppered glass vials securable with an aluminium crisp seal, which are labelled accordingly. Also included are 50ml of dilution buffer, 50ml of elution buffer, 14ml of lyophilised substrate, 14ml of lyophilised oxidising agent and 4 x 3.5ml of lyophilised coenzyme. This size of kit enables the researcher to carry out approximately 500 tests.
EXAMPLES
The Kit described above was used to demonstrate the effectiveness of the second reading on a Cobas Mira Plus ( Roche Diagnostics ). The experiment was set up to mimic the effect of incomplete elution on the measurement of enzyme activity. According to the assay protocol described above 2 spots of 3.2mm must be punched from the sample and inserted into the sample cup. This was done using whole blood samples of 3 human donors spotted and dried onto filtration paper. The instrument was programmed for the second reading by repeating the assay from the same sample, only with a single reading at 405nm instead of the kinetic reading at 340nm. To mimic the effect of incomplete elution the same samples were used but only one spot was punched and inserted into the sample cup. The results are shown below in Table 1 : Table 1 mOD/ min 405nm
340nm
Units 2 spot 1 spot °/o 1 spot 2 % spot
MMRC02 N 19.42 9.98 51.4% 1.81 3.28 55.3%
MMRC02 N 23.18 11.2 48.3% 1.92 3.81 50.5%
USA 1 8.29 4.2 50.7% 1.04 2.07 50.2%
USA 1 7.76 4.71 60.7% 1.12 1.97 56.8%
USA 2 12.98 5.98 46.1% 1.16 2.48 46.9%
USA 2 13.05 6.07 46.5% 1.14 2.40 47.7%
Ratio Norm
alised
2 spot 1 spot 1 spot 2 spot %
MMRC02 N 1.000 1.810 18.06 19.42 93.0 %
MMRC02 N 1.162 1.705 19.10 19.96 95.7 %
USA 1 0.631 3.154 13.25 13.14 100. 8%
USA 1 0.601 2.932 13.81 12.92 106. 9%
USA 2 0.756 2.823 16.88 17.17 98.3
%
USA 2 0.732 2.866 17.40 17.84 97.6 %
It can be : seen that :
1) The result of measurement at 405nm is proportional to the amount of blood eluted from the sample.
2) The results can be normalised by relating the 405nm results to a reference sample
3) Any wavelength in the range 420-400nm or 535-545nm or 565-580nm will work in the same aspect.
Verifying an alternative wavelength for Hb.
It has been established that the optimum wavelength to measure the Hb content is at 414 or 417nm. However, this would limit the usefulness of the proposed assay to the labs using automated analyzers or spectrophotometers because the microplate readers have a limited range of optical filters available.
One such filter being the 405nm optical filter. It was examined whether this particular wavelength, though not optimal, is Hb specific and yields a high enough signal. The first thing was to make sure that the rest of the chemicals did not absorb in the 380-460 area nor do the products of the kinetic reaction.
Firstly the spectrum of a normally lyzed whole blood sample, using the Elution
Buffer contained in the kit was obtained, before and after the addition of the
Reagent Mixture and the first spectrum was subtracted from the second (Figure 1). It is clear that the reagents absorb only at wavelengths below 360nm. The elution buffer gave an almost identical spectrum to deionized water (data not shown here).
Dried Blood Soot Samples. The results shown in Figure 3 demonstrate the suitability of the 405nm optical filter for an accurate determination of the Hemoglobin content of the samples. The dried blood spots are far more difficult to be evaluated because the blood actually eluted greatly varies from sample to sample. The same sample may give a 10% difference when examined under the same experimental conditions. On top of that, the blood initially spotted on each disk is not the same. The situation is further complicated in the "punching" stage where not everyone punches at the center of the spot or, even worse, not all the disk cut is covered with blood. Thus, apart from the normal differences between samples which are due to parameters like hematocrite or average Hb content per erythrocyte, many other parameters that have to do with the assay itself magnify them. This is clearly shown in Figure 3. Twenty five
dried blood spot samples from a randomly selected microplate are shown in this Figure. It can be seen that the Hb content of the spots varies from 0.076 at 405nm (0.107 at 417nm) to 0,528 at 405nm (0.731 at 417nm). The average OD at 405nm is 0.307 ± 0,126 (0.442 ± 0.179 at 417nm). However, the ratio of the ODs (405nm / 417nm) is amazingly stable (light gray bars) under these varying conditions (0.693 ± 0.031). A total of 2.000 neonates have been examined and OD values ranging from 0.040 to 0.610 at 405nm have been found. However the average ratio (n=2000) was almost identical to the one mentioned earlier (0.701 ± 0.0309). This is an indication that the reading at 405nm can substitute the reading at the optimal wavelength (414nm or 417nm).
Whole Blood samples Results from 25 whole blood samples (a randomly selected microplate) are shown in Figure 4. The ODs ranged from 0.735 at 405nm (0.837 at 417nm) to 1,115 at 405nm (1.351 at 417nm). In this process, only pipetting inconsistency may alter the Hb content of each sample. The average OD at 405nm (417nm) for these 25 samples was 0.962 ± 0.103 (1.159 ± 0.156), while the ratio of ODs (405nm / 417nm) was found 0.834 + 0.038. Even when whole blood was further diluted 1 :2 and 1 :4 with normal saline, the ODs observed was proportionally lower but the ratio of the ODs (405nm / 417nm) remained the same (0.822 + 0.037 for an 1 :2 dilution, n = 25; 0.819 ± 0.038 for an 1 :4 dilution, n=30).
Normalization of Samples for Hb content. The activity of the G-6-PD enzyme in this assay is measured indirectly by the formation of β-NADPH following the oxidation of glucose-6-phosphate to 6-phosphogluconate. The NADPH produced is measured at 340nm in kinetic mode and the change of OD per minute is proportional to the G-6-PD activity of the sample. By definition, one International Unit of G-6-PD activity is the amount of G-6-PD activity which will convert lmmol of substrate per minute under the conditions (mainly temperature) specified in the assay. This activity is directly proportional to the number of erythrocytes actually lyzed during the elution process which can be expressed as grams of Hemoglobin in the sample. In order to have comparable results, the activity measured in each sample has to be calculated for the same amount of Hb irrespectively of the actual Hb contained in that sample. Since all comparisons are made against the controls, the Hb of the controls should become the basis of comparisons. We therefore divided all ODs measured at 405nm following the kinetic (ODsamPie405nm) by the OD at 405nm of the control (ODControi405nm)- The resulting values represent the "relative" Hb content of each sample as compared to the control. The activity measured in each sample (ΔODsampιe34onm/min) is then divided by the "relative Hb" factor in order to get the activity of the sample which corresponds to the Hb content of the control.
Expressing Results in Units / gram Hb. The G-6-PD activity of the sample is now directly comparable to that of the control (δODcontroi34onm/min). Since the control's activity is already given in Units / g Hb (U/g Hb) the sample's activity can be readily expressed in the same units. Special care should be taken to use the controls which are rated at the temperature used for the assay. Thus, when using analyzers or microplate " readers equipped with a heating device, any control can be used. When performing the assay at room temperature it is preferable to use controls which are rated at 24°C. The equation that gives results expressed in U/g Hb is given below.
δODSample340nm/inil / δODcontrol340nm/ min X Control Value = Sample Activity in U/g Hb
ODsamp|e405nm / ODcontro|405nm
Results for verification. In order to verify that this equation will give accurate and reproducible results the following experiment was performed. A set of ten samples was prepared as shown in Table 2. Thirty sets of these samples were tested in the same day using the Single Test kit in order to simulate a multiple day screening condition (different Reagent Mixture for each set). One box (MMR010, Batch 01/98) was opened and used for each set.
Table 2 'Indicates Applicant's samples
Neonate #1 was classified as "Normal" during screening whereas Neonate #2 and #3 were classified as "deficient" (using the currently employed semi- quantitative Beutler method). All samples were compared to * Normal in column 2 and the assay was performed at 37°C. These results are presented in Table 3.
Table 3
It is evident that this method has a very good accuracy and reproducibility and the results obtained from different sessions and/or days are directly comparable to each other. Limitations of the method.
This method assumes that the optical properties of the samples are the same with the controls. Since the molecular extinction coefficient of the Beer Lambert Law (e) applies for a given compound this method can be used to directly express results in U/g Hb by comparing them to controls tested with the same
Reagent Mixture and under the same conditions. Next day's microplates should be compared against controls tested with the Reagent Mixture used at that day. However results expressed in U/g Hb are directly comparable irrespectively of when they were obtained.
Workable Wavelengths :
In order to determine which wavelengths could be used for normalisation of Hb content we run the same samples 100 times. Average values were then inserted in tables 4 and 5 in order to determine which wavelengths satisfied the criteria
(less than 5% fluctuation from the mean ratio to the optimal wavelength at
414nm).
Table 4
Hb NORMALIZATION PROCEDURE
Reference Value OD at 414 nm (n=300)
WAVELENGTHS TESTED { m) ~ RATIOS (O0X
S.WL DeSCRIPTIUN dilution 460 RATT 450' RATIO 440 RATIO Whole blood (March 98) 1 : 1 dil 0,352 0,219 0,336 0,21 0,423 0,264
6
Whole blood (March 98) 1 :2.5 dil 0,142 0,208 0,118 0,173 0,103 0,151
2
Sigma Normal control none 0,112 0,208 0,083 0,155 0,059 0,11
6
Whole blood (Nov.97) 1 : 1 dil 0,191 0,174 0,188 0,171 0,217 0,198 Whole blood (Nov.97) 1 : 1 dil 0,154 0,154 0,151 0,152 0,403 0,405
Peak observed at 415,5 nm in all cases
Acceptance criteria : % variance less then 5% for n = 100 samples at least
CONCLUSION : RANGE OF ACCEPTABLE ODs (420 400 nm) Reference values:
414 1,603 0,682 1,17 0,537 1,098 0,994
Table 5 - Results of OD 405nm / 417nm whoie/417 whole/405 W405/417 spot/417 spot/405 S405/417 whole/417 whole/405
1 131 0.937 0 8284704 0.384 0.249 0 6484375 1.131 0 937
1.216 0.972 0 7993421 0.625 0 437 0.6992 1.216 0 972
1.2 0.963 0.8025 0.455 0.294 0 64615385 1.2 0.963
1.137 0.926 0.8144239 0.482 0.343 0,71161826 1.137 0.926
0.84 0.799 0.9511905 0.539 0.355 0.65862709 0.84 0.799
1.366 1.073 0.7855051 0.568 0.402 0. 70774648 1.366 1.073
1.273 1.013 0 7957581 0.453 0.302 0.66666667 1.273 1 013
1.223 0.977 0.7988553 0.557 0.386 0.6929982 1.223 0 977
1.276 1.036 0.8119122 0.504 0.341 0.6765873 1.276 1.036
1.145
1.411 1.145 0.8114812 0.731 0.528 0.72229822 1.411
1.18 0.959 0.8127119 0.53 0.34 0.64150943 1.18 0.959
1.215 1.035 0 8518519 0.635 0.435 0.68503937 1.215 1.035
0 973 0.839 0 8622816 0.478 0.319 0.66736402 0.973 0 839
1.246 1.088 0 8731942 0.537 0 38 0.70763501 1 246 1 088
1.054 0 9 0 8538899 0.465 0.327 0 70322581 1.054 0 9
1.206 0.999 0,8283582 0.551 0.389 0 70598911 1.206 0 999
1.077 0.928 0 8616527 0.528 0.375 0.71022727 1.077 0 928
1.088 0.934 0.8584559 0.47 0.332 0.70638298 1.088 0 934
1.309 1.038 0 7929717 0.487 0.35 0 71868583 1.309 1 038
1.351 1.115 0.8253146 0.609 0 429 0.7044335 1.351
1.115
1.109 0.906 0.8169522 0.107 0.076 0 71028037 1.109 0 906
0.837 0.735 0 8781362 0 131 0.093 0 70992366 0.837
0.735
0.929 0.806 0 8675996 0.177 0.131 0 74011299 0.929 0 806
0.514 0 405 0.78793774 0.38 0 2 9
0.111 0 077 0 69369369 0.625 0 37
0.179 0.122 0.68156425 0 455 0 294
0.119 0.083 0.69747899 0.482 0 343 0.539 0 355
0.961871.158565 0.834035 0.307407 0.441704 0.6963636 average 0.568 0 402
0.1032170.155838 0.038382 0.126856 0.179513 0.030744 st.dev. 0.453 0.302 0.557 0.386 dιl.blood/405 dιl.blood/417 405/417 0.504 0 341
0.428 0.545 D.7853211 0.731
0.528
0.481 0.587 0.8194208 0.53 0 3
0.4545 0.566 0.802371 average 0.635 0 435
0.037477 0.029698 0.024112 st.dev. 0.478 0 319 0.537 0 38 0.465 0 327 0.551 0 389 0 528 0 375
0.47 0.332
0.487 0.35
0.609 0.429
0.107
0.076
0.131 0.093
0.177 0.131
0.514 0.405
0.111 0.077
0.179 0.122
0.119 0.08
It will be apparent to those skilled in the art that there are many variants of the method described herein before. For example, in some cases where a colour reagent is added at the end of the assay obtaining the necessary measurements at the wavelength needed are prohibited or the reaction produces a final product which also absorbs at 410nm. In these cases it is necessary to instead of measuring λ2 after the kinetic (ODCλ2 or ODSλ2) it can be measured before or after addition of the chemicals, colour reagents or even before running the kinetic measurements. The advantages of the method of the present invention are as follows:
1) It provides a QUANTITATIVE measurement of enzyme activity due to the Kinetic Readings. Quantitative results are needed to differentiate between intermediate and deficient specimens. It is very difficult to do this "by eye" and is also subjective. 2) There is no interference from other enzymes due to the use of a Pre- Incubation. Other enzymes can interfere with the reaction as has been described by Solem et. al. These enzymes react with the coenzyme to form the same end-product as with the enzyme to be measured. The solution to this is the use of a pre-incubation during which the interfering enzyme(s) are active until completion of their reaction. After this the enzyme to be measured solely uses the coenzyme to produce the measured end-product, which is called the linear phase. The invention makes use of this linear phase for the kinetic readings.
3) The reaction can take place at a wide range of temperatures 22-37C. Using a wide range of temperatures allows the laboratory to choose a method best fit for the routine workflow. Working at room temperature means there is no need for temperature controlled incubators.
4) It provides a truly QUANTITATIVE result by assessment of the haemoglobin content of the sample and calculation of the results in Units per gram Haemoglobin. This is an absolute necessity if one is to measure enzyme activity. Enzymes are often located inside the red blood cells and bound to haemoglobin. The amount of enzyme activity measured in a sample is directly proportional to the amount of haemoglobin. This must be measured in the same eluate used for the reaction. Elution of dried bloodspots can be incomplete and is always subject to variation. So far it has not been possible to asses the Hb content of an eluate.