Optimised MRI acquisition protocols for Siemens 3 T and 7 T systems running XA60.
π΄ Active development: This repository may change without notice. Consider it complete only when this banner has been removed.
- Protocol collections
- Motivation
- Design principles
- What this repository contains
- Recommended workflow after data acquisition
- bto-docker
- Help and issue reporting
- Use in research
- Relevant publications
- Contributions
- Previous version
- Contact
Each page describes its collection and will link to the matching Zenodo package. The 3 T collection includes anatomy, fMRI, diffusion, and ASL. The 7 T collection includes anatomy, fMRI, and diffusion.
The 7 T collection does not include a product ASL protocol. To discuss 7 T ASL options, reach out via email (see end of page).
Prot C9AA ocol fragmentation is a quiet crisis in neuroimaging research. Standardisation efforts are not keeping pace with the need, especially in clinical research. Scanner time and research funding are also under increasing pressure, while data are being reused more often. A control group acquired for one study may support another study, and datasets collected at different sites may be combined.
That depends on acquisition parameters being shared and understood. In practice, protocols often vary between studies at the same site. Small changes in resolution, timing, acceleration, reconstruction, or coil configuration can make data harder to compare and can affect later processing. The problem is larger in multi-site studies, where local scanner setups and established practices already differ.
Access is another constraint. Hospitals and research groups may not be able to obtain or maintain specialised research sequences. BRAIN-TO uses Siemens product sequences so that collaborating sites can implement the protocols without Work-in-Progress packages or C2Ps.
BRAIN-TO provides a stable starting point for researchers, MR physicists, and technologists who need dependable protocols without designing every sequence from scratch. The aim is consistent performance and practical reuse, rather than pushing individual scanner settings to their limit.
- Siemens product sequences make the protocol sets practical to share across sites.
- Isotropic voxels support consistent spatial sampling and reduce partial-voluming concerns.
- Most scans target 5 to 6 minutes at 3 T and 6 to 8 minutes at 7 T (exceptions exist).
- Coil-specific variants use the acceleration available from the head coil.
- Data acquired with BRAIN-TO protocols have been evaluated with community-standard MRI processing software.
- Protocol index and compatibility reference
- 3 T helper scripts
- 7 T helper scripts
- GitHub issue forms for protocol, acquisition, and processing support
Zenodo is the single source for downloadable protocol packages and their full acquisition parameters. The repository provides an index, practical guidance, and links to related tools.
- Use dichotomise to check, sort, rename, de-identify, and archive DICOM exports from Siemens XA60+ systems.
- Convert the organised data to BIDS with BIDScoin.
- Use bto-docker, a general-purpose BTO protocols processing container with FSL, oxasl, FreeSurfer, and supporting analysis scripts for ASL perfusion and structural MRI.
Each repository documents its supported inputs, versions, and release locations. Check its README before using it with a BRAIN-TO dataset.
General-purpose processing container for BTO protocols.
| Tool | Version |
|---|---|
| FSL | 6.0.7.22 |
| FreeSurfer | 8.2.0 |
| oxasl | custom |
| ANTs | 2.6.5 |
| AFNI | latest |
| MRtrix3 | latest |
docker pull brainto/bto-docker:latestMount your data directory to /data:
docker run --rm \
--user $(id -u):$(id -g) \
--cpuset-cpus "0-9" \
-v /path/to/data:/data \
brainto/bto-docker \
bto-asl --helpFreeSurfer requires a valid license file mounted at runtime:
docker run --rm \
-v /path/to/data:/data \
-v /path/to/license.txt:/opt/fs-license/license.txt \
brainto/bto-docker \
bto-anat freesurfer --helpAll file paths must be absolute paths as seen from inside the container (i.e. under /data).
| Command | Purpose |
|---|---|
bto-asl |
ASL perfusion pipeline |
bto-anat fsl |
FSL structural preprocessing |
End-to-end pipeline.
docker run --rm -it \
--user $(id -u):$(id -g) \
--cpuset-cpus "0-9" \
-v $PWD:/data \
brainto/bto-docker bto-asl \
--subjid sub-01_ses-rest \
--asl /data/sub-01/ses-rest/perf/sub-01_ses-rest_asl.nii.gz \
--t1 /data/sub-01/ses-anat/anat/sub-01_T1w.nii.gz \
--m0-fwd /data/sub-01/ses-rest/fmap/sub-01_ses-rest_dir-PA_m0scan.nii.gz \
--m0-rev /data/sub-01/ses-rest/fmap/sub-01_ses-rest_dir-AP_m0scan.nii.gzJSON sidecars (.json alongside each NIfTI) are auto-detected for acquisition parameters. Pass --no-json to skip sidecar detection and use assumed PE direction.
If structural preprocessing has already been run, pass --anatdir to reuse the existing .anat directory:
--anatdir /data/sub-01/ses-anat/anat/sub-01_T1w.anat| Flag | Default | Description |
|---|---|---|
--subjid |
required | Subject ID β names the output directory bto-asl_<subjid>/ |
--asl |
required | 4D ASL NIfTI (repeat for multiple runs) |
--t1 |
β | T1w structural NIfTI β triggers FSL structural preprocessing |
--anatdir |
β | Existing .anat directory β skips structural preprocessing |
--m0-fwd |
β | Forward-PE M0 calibration image |
--m0-rev |
β | Reverse-PE M0 for topup distortion correction |
--outdir |
alongside --asl |
Base output directory |
--no-json |
off | Skip JSON sidecar requirement |
--pe-dir |
j |
Forward PE direction when --no-json is active |
docker run --rm -v /path/to/data:/data \
brainto/bto-docker bto-anat fsl \
--t1 /data/sub-01/anat/T1w.nii.gzRuns bias correction, brain extraction (mri_synthstrip), linear and nonlinear MNI registration, FAST tissue segmentation, and subcortical segmentation. Output defaults to <t1_parent>/derivatives/anat/<t1_stem>.anat/.
- FreeSurfer license is never baked into the image β mount it at runtime to
/opt/fs-license/license.txt - Base image:
ubuntu:24.04 - All scripts use FSL's conda Python (
/usr/local/fsl/bin/python3) as the runtime interpreter - Default Docker runs as root; pass
--user $(id -u):$(id -g)to write output as the current user
Use the issue form that matches the problem:
Before using a protocol, confirm that it is appropriate for the local scanner configuration, coil, participant group, safety procedures, ethics approval, and study question. Local MR physics and technologist review remains essential.
- Kashyap S, UludaΔ K. Advancing Clinical and Neuroscientific Research Through Accessible and Optimized Protocol Design at 3T. Siemens MAGNETOM Flash, RSNA Edition, 2023. Read the article
- Kashyap, S. (2026). Advancing clinical neuroscience research through accessible, optimised protocol design at 3 and 7 T. Zenodo. Annual Meeting of the Organization for Human Brain Mapping (OHBM), Marseille, France. https://doi.org/10.5281/zenodo.22769256
Contributions are welcome. Useful contributions include protocol feedback, implementation experience at another site, field-specific scripts, processing pipelines, and documentation improvements. Please open an issue before proposing a substantial change.
The complete XA30 protocol set is available on the
xa30_archive branch.
For questions about protocol selection, implementation, or 7 T ASL, email sriranga.kashyap@utoronto.ca.